| Model Number |
191128 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility biomedical technician (biomed) contacted fresenius technical support (ts) to request a replacement blood pump module for a 2008t hemodialysis (hd) machine.It was reported that the blood pump rotor had cut into the blood tubing segment during a patient¿s treatment, resulting in approximately 2 ml of blood loss.At first, an air detector alarm sounded, and the staff then noticed air in the blood lines.The blood pump was stopped, and the blood tubing was noted to be damaged.The patient was moved to another machine where they completed their treatment.The biomed told ts that the event did not result in any patient harm or injury.Multiple attempts to obtain additional information have been performed without success.Thus far, no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3 plant investigation: the blood pump module was returned to the manufacturer for physical evaluation.The blood pump was returned without the rotor or the tubing retaining clamp.During visual examination of the returned part, no physical damage was found.A ¿known good¿ rotor and a tubing retaining clamp were installed onto the blood pump for testing.Next, a ¿known-good¿ blood line with dialyzer was inserted into the blood pump module and was submerged in water for testing.There were no problems during the initial power up.The blood pump module functioned properly during dialysis without any alarms or failures.The blood line did not leak during testing and no punctures were found on the blood line after use.The self-test program was completed without failures.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The device performed as designed and an associated cause could not be determined.
|
| |
|