ZIMMER BIOMET, INC. STRAIGHT SHELL INSERTER; HIP INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial right hip arthroplasty, the surgeon went to insert the shell into the patient and the tip of the screw threads on the inserter broke off in the shell.Cup and inserter were removed and a new shell was used to complete the surgery.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event is unable to be confirmed due to limited information provided by the customer.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
H6: type of investigation ¿ remove code 4114 device not returned.Complaint sample was returned and evaluated against the reported event.Visual evaluation of the product found the lead threads of the hex bolt are fractured and missing.There was wear and tear consistent with use.No other damage is noted.Evaluation of the returned product does not change the outcome of the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SHELL INSERTER
Type of Device
HIP INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11087092
MDR Text Key224133019
Report Number0001822565-2020-04242
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00780401520
Device Lot Number61970607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNK SHELL.; UNK SHELL
-
-