Model Number 71119105
Device Problems Corroded (1131); Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during surgery, the 7.0 cann sc measure gauge presented rust and is difficult to read the numbers.Procedure continued with the same device, without a delay or an injury.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to have some discoloration, and the words on the device were not legible, rendering the device inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.0 CANN SC MEASURE GAUGE
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11087478
MDR Text Key224244160
Report Number1020279-2020-07879
Device Sequence Number1
Product Code HRS
UDI-Device Identifier03596010428363
UDI-Public03596010428363
Combination Product (y/n)N
PMA/PMN Number
K123055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71119105
Device Catalogue Number71119105
Device Lot Number16JM10036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Date Manufacturer Received06/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-