• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 5.5MM 45, HD BARIATRIC AUTOCLAVABLE LAPAROSCOPE 45CM; SYSTEM, X-RAY, MAMMOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 5.5MM 45, HD BARIATRIC AUTOCLAVABLE LAPAROSCOPE 45CM; SYSTEM, X-RAY, MAMMOGRAPHIC Back to Search Results
Model Number 0502513045
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was a blurry image.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure [cloudy] was confirmed.According to henke: scope evaluated as a level 3.Distal tip and fiber damage, loose optics, bent needle, scratches and residue.The complaint for ¿cloudy lens¿ has been confirmed.Probably root cause for this failure is: due to customer use and handling.The manufacture date is not known.
 
Event or Problem Description
It was reported that there was a blurry image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IDEAL EYES 5.5MM 45, HD BARIATRIC AUTOCLAVABLE LAPAROSCOPE 45CM
Common Device Name
SYSTEM, X-RAY, MAMMOGRAPHIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11088011
Report Number0002936485-2020-00561
Device Sequence Number3859223
Product Code IZH
UDI-Device Identifier07613327059281
UDI-Public07613327059281
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K993045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502513045
Device Catalogue Number0502513045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2020
Initial Date Received by Manufacturer 12/04/2020
Supplement Date Received by Manufacturer12/04/2020
Initial Report FDA Received Date12/29/2020
Supplement Report FDA Received Date02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-