BAXTER HEALTHCARE CORPORATION EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400L
Device Problem Material Twisted/Bent (2981)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.A device history review revealed no issues that could have caused or contributed to the reported issue.A visual inspection was performed which identified bent legs.The reported condition was verified.The cause of the condition was related to component failure; most probable due to end user handling.All four feet were replaced to correct the condition.The device was serviced to meet functional specification.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the leg of an em 2400, load cell module device was bent and would not fit back onto the base.This was identified during an unspecified process step.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key11088366
MDR Text Key224764112
Report Number1416980-2020-08138
Device Sequence Number1
Product Code NEP
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-