• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION INFINION 16; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION INFINION 16; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-2316-50E
Device Problem Low impedance (2285)
Patient Problems Cerebrospinal Fluid Leakage (1772); Headache (1880); Therapeutic Effects, Unexpected (2099); Discomfort (2330); No Code Available (3191)
Event Date 12/03/2020
Event Type  Injury  
Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.Additional suspect medical device component involved in the event: product family: scs-linear leads, upn: (b)(4), model: sc-2316-50e, serial: (b)(4), batch: 7108294.
 
Event Description
It was reported that the patient had a dural puncture during the procedure.After the procedure, the patient was experiencing spinal headache on the right side of her head and the impedances showed very low.The device was turned on and the stimulation was increased however it caused a tightening feeling and discomfort.The physician decided to pull the leads and the patient was given a caffeinated beverage.The headache got a little better and the explanted leads were discarded by the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINION 16
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
RQ   00646-2602
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key11088381
MDR Text Key224094967
Report Number3006630150-2020-06338
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729797807
UDI-Public08714729797807
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/19/2022
Device Model NumberSC-2316-50E
Device Catalogue NumberSC-2316-50E
Device Lot Number7108126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
-
-