| Catalog Number |
2400M |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the green pump door of an exactamix automated compounding device has missing parts.It was further reported the green door had missing screws.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.A visual inspection was performed which noted the device was missing the screws from the green door.Functional and electrical testing was performed with no issues noted.The reported condition was verified.The cause of the missing screws was likely from normal everyday use.To correct the condition, the door and screws will be replaced during refurbishment of main module.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|