| Model Number |
100149282 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Output Problem (3005)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
During the atrial fibrillation ablation procedure, there was a shift on the mapping system and an error message on the ampere, and the procedure was cancelled.After some ablation had been completed, the tip of the catheter appeared to fly out of the geometry on ensite, and the error message " 'amplifier error.Log out and power cycle both velocity dws and amplifier before restarting.Contact sjm technical support." was displayed.The case was cancelled, and the patient experienced no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.A review of the dhr was performed and ruled out the possibility of a manufacturing related issue causing or contributing to the reported complaint issue.Based on the information received, the cause of the reported incident could not be determined.
|
| |
|
Search Alerts/Recalls
|