BIOTRONIK SE & CO. KG COROX OTW 85-BP; LV LEAD Back to Search Results
Model Number 354807
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 12/28/2020
Event Type  Injury  
Event Description
This system was explanted due to suspected infection.Should additional information become available, this file will be updated.
 
Manufacturer Narrative
An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COROX OTW 85-BP
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key11091954
MDR Text Key224245405
Report Number1028232-2020-05703
Device Sequence Number1
Product Code OJX
UDI-Device Identifier04035479093549
UDI-Public04035479093549
Combination Product (y/n)N
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2014
Device Model Number354807
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/28/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received12/30/2020
Supplement Dates FDA Received12/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age53 YR
-
-