• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24672
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 4.0 x 150, 135cm mustang balloon catheter was advanced for dilatation.However, during initial inflation, the balloon pressure could not maintain.It was noted that the pressure was 4 atmospheres and lasted for 5 seconds.When the device was removed, it was noticed that the balloon was ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 4.0 x 150, 135cm mustang balloon catheter was advanced for dilatation.However, during initial inflation, the balloon pressure could not maintain.It was noted that the pressure was 4 atmospheres and lasted for 5 seconds.When the device was removed, it was noticed that the balloon was ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: mustang device - 4x15x135 was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.A microscopic examination identified a balloon pinhole located approximately 14mm proximal of the distal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.No other issues were identified during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11092204
Report Number2134265-2020-18563
Device Sequence Number16827274
Product Code LIT
UDI-Device Identifier08714729793694
UDI-Public08714729793694
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2023
Device Model Number24672
Device Catalogue Number24672
Device Lot Number0025524763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer01/25/2021
Initial Report FDA Received Date12/29/2020
Supplement Report FDA Received Date02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-