| Model Number |
PS10.0-80D265 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.Root cause is unable to be determined at this time.The reported event has been addressed in the device labeling.
|
| |
|
Event or Problem Description
|
|
Information was received via literature that the patient had disseminated cortical osteolyses only on the right side level with the nail¿s telescopic junction which was clearly atypical and potentially implant-related.Complete resolution of the osseous alterations was achieved.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number for the device is unknown.However, two potential lot numbers were provided: 0013119aaa (manufacture date of jan 31, 2020).0013123aaa (manufacture date of jan 31, 2020).Journal article citation: frommer, adrien, et al."focal osteolysis and corrosion at the junction of precice stryde intramedullary lengthening device." bone and joint research, vol.10, no.7, 16 july 2021.Https://online.Boneandjoint.Org.Uk/doi/epub/10.1302/2046-3758.107.Bjr-2021-0146.R1.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the patient had lytic osteolysis and that they reported pain during distraction and consolidation.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data.
|
| |
|