• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. GEM V/NV 20DP CKV 2SS 117-IN 20PK; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. GEM V/NV 20DP CKV 2SS 117-IN 20PK; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 2426-0500
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6) hospital.Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.(b)(4).
 
Event Description
It was reported that the gem v/nv 20dp ckv 2ss 117-in 20pk experienced tubing expansion/ballooning.The following information was provided by the initial reporter: complaint 2 of 2.Material #: 2426-0500, lot#: unknown.Rn reported that the tubing herniated when the start button was pushed on the alaris pump when starting a new infusion.The rn saw it immediately and was able to stop the infusion and able to set up a new tubing set.
 
Event Description
It was reported that the gem v/nv 20dp ckv 2ss 117-in 20pk experienced tubing expansion/ballooning.The following information was provided by the initial reporter: complaint 2 of 2.Material#: 2426-0500, lot#: unknown.Rn reported that the tubing herniated when the start button was pushed on the alaris pump when starting a new infusion.The rn saw it immediately and was able to stop the infusion and able to set up a new tubing set.
 
Manufacturer Narrative
Investigation summary: one sample was received for quality investigation.The customer complaint that the tubing herniating was confirmed by visual inspection.The silicone tubing was noticeably deformed in shape from the failure that the customer reported.No further damages or defects were observed.A device history record review could not be performed on model 2426-0500 because a lot number was not provided by the customer.The root cause of this failure can be attributed to a failure of the component.The sample was able to prime and infuse without issues, and therefore it is determined that the silicone tubing failed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEM V/NV 20DP CKV 2SS 117-IN 20PK
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana 22244
MX  22244
MDR Report Key11092372
MDR Text Key230054865
Report Number9616066-2020-20681
Device Sequence Number1
Product Code FPA
UDI-Device Identifier07613203020992
UDI-Public07613203020992
Combination Product (y/n)N
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2426-0500
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Date Manufacturer Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-