| Catalog Number |
MP5305-C |
| Medical Device Problem Codes |
Complete Blockage (1094); Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The initial reporter also notified the fda on (date) via medwatch # mw5096890.(b)(4).Investigation summary: one sample was returned for investigation.An investigation was performed.Sample was connected to a bd syringe and attempted to flush water through the set.Set was able to be primed.The customer complaint that the maxplus extension set would not flush after multiple attempts could not be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review for model mp5305-c, lot number 20055016 was performed.The search showed that a total of (b)(4).There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the 7-in pressure rated set w/ maxplus was clogged/blocked/occluded.The following information was provided by the initial reporter: complaint 2 of 2: material #: mp5305-c , batch/ lot #: 20055016.Multiple attempts were made to flush through buff cap on end of extension set (with various from same lot number) and were unable.Had to change buff cap and able to flush and infuse ivf.
|
| |
|