Model Number MMT-1715K |
Device Problems
Excess Flow or Over-Infusion (1311); Device Difficult to Program or Calibrate (1496); Loss of Data (2903); Moisture or Humidity Problem (2986)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/23/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Information received by medtronic indicated that the insulin pump had a stuck button alarm and pump error alarm.Customer also reported insulin was came out of reservoir and insulin pump was over delivering the insulin.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.The reservoir will not be returned for analysis.
|
|
Manufacturer Narrative
|
Device was programmed with multiple boluses and monitored.All boluses delivered properly and were listed in the daily history screen.Device then was programmed with multiple basal profiles and monitored.All basal profiles delivered their indicated amounts and were verified in the daily and summary history screens.The units left at the pump display matched properly the units left at test reservoir.No over delivery anomaly, bolus anomaly or basal anomaly noted during testing.No keypad unresponsive or exposed to moisture noted during testing.(b)(4).
|
|
Search Alerts/Recalls
|