Device received with blank solid white display after battery insertion due to severe corrosion on electronic board stack.Unable to verify critical pump errors alarms, charge/battery lasts less than expected anomalies or low reservoir anomalies due to solid white display anomaly.Unable to perform displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self test due to solid white display anomaly.Corroded battery tube.Corrosion on force sensor assembly and moisture damage on motor assembly.Device received with pillowing keypad overlay, faded/stained/peeling serial number label, peeling/fading end cap address label and scratched case.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|