Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test at 0.0876 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.No drive/motor anomaly noted.The motor was tested outside of device and passed.No cosmetic damage noted.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|