• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I FULL OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 8.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I FULL OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number IFNT585
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); Unspecified Infection (1930); Inflammation (1932); No Code Available (3191)
Date of Event 08/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary reporting.
 
Event or Problem Description
It was reported that clinician observed peri-implantitis.The patient was implanted on (b)(6) 2019, and there was no adverse reaction after the implantation.After 3 months, the patient didn¿t feel well.After examination, it was found that there was inflammation around the implant.The implant was removed.Edema.Tooth location 6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key11094408
Report Number0001038806-2020-02163
Device Sequence Number767494
Product Code DZE
UDI-Device Identifier00844868005094
UDI-Public(01)00844868005094(17)211110(10)2016102050(241)IFNT585
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/10/2021
Device Model NumberIFNT585
Device Catalogue NumberIFNT585
Device Lot Number2016102050
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2020
Initial Report FDA Received Date12/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-