• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ROTATABLE CLIP FIXING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ROTATABLE CLIP FIXING DEVICE Back to Search Results
Model Number HX-201UR-135L
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Laceration(s) (1946)
Type of Reportable Event Serious Injury
Event or Problem Description
It is reported in user facility med watch ((b)(4)) and follow-up information provided by the customer, during an emergent repair of a gastrointestinal (gi) bleed, a single use rotatable clip fixing device failed to detach when fired (reported in case with patient identifier (b)(6)).This caused the tissue to tear and increased bleeding to occur.A second clip was deployed, and it also failed to detach appropriately (reported in case with patient identifier (b)(6)).Two other clips were test fired into a blanket and those failed to detach appropriately (this report, and case with patient identifier (b)(6)).Additional details and series of events as follows: at the beginning of the procedure, irrigation was performed, removing a clot.This revealed a slow pulsating bleed.Attempted to control bleeding with clip- clipping device malfunctioned.One clip fired but did not detach from the device.This caused the tissue to tear and increased bleeding to occur.The surgeon attempted to place another clip, but the same thing happened.At that point the patient started to bleed briskly.A second surgeon was called in to assist.The second surgeon did not attempt to place any more clips because he did not want to lose control of the vessel again the surgeons attempted to try a heat probe to control the bleeding without success.The area was ultimately injected with epinephrine until the bleeding was controlled.Hemospray was then applied on top.It was determined that the epinephrine and hemospray held to allow enough time to get patient transferred out to a different facility since patient still needed a more permanent method to control the bleeding.The patient was transferred to another facility for a higher level of care.Follow-up information received from subsequent facility.They performed urgent endoscopy at the bedside.One previously placed clip was noted at the base of the ulcer adjacent to the visible vessel.They attempted four times to place clips (unspecified manufacturer) at the base of the visible vessels, only one clip attached, and the other 3 clips spontaneously dislodged after application.The necrotic material on the surface of the visible vessel sloughed off as the clips dislodged after deployment.The clips were not holding as there was no mucosa for them to get an anchor to hold on.Procedure was paused to give the patient a break, then resumed the endoscopy.The physicians noted two dislodged clips which had moved back to the stomach.To prevent aspiration, the two clips were grasped with forceps and moved back to the duodenum.The scope was advanced and again the physicians noted active pulsatile bleeding from the visible vessel.The decision was made at that time to take the patient to emergent surgery because the active arterial bleeding was not amenable to endoscopic control.Interventional radiology (ir) was contacted, and embolization was determined to be the appropriate intervention.The patient's current condition could not be provided by the customer (since they were transferred out of their facility).The complaint devices will not be returned to olympus for physical evaluation.They were discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ROTATABLE CLIP FIXING DEVICE
Common Device Name
SINGLE USE ROTATABLE CLIP FIXING DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11094820
Report Number2951238-2020-00532
Device Sequence Number824217
Product Code PKL
UDI-Device Identifier04953170353161
UDI-Public04953170353161
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 12/30/2020
Report Date (Section F) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHX-201UR-135L
Device Lot Number98V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/03/2020
Event Location Hospital
Date Report to Manufacturer12/03/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-