SMITH & NEPHEW, INC. JOURNEY RESECTION CHECK; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74012431
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during tka procedure the angel wing was getting stuck when checking journey resection check through the cut block (inside the patient).Used a saw blade instead and was able to complete procedure without delay or harm.Post procedure, i compared angel wing with another angel wing (same part number) and it was confirmed that the angel wing in question had been warped.The procedure was finished with a change in surgical technique without surgical delays.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.According to clinical/medical investigation, the ¿angel wing was getting stuck when checking¿ the resection through the cut block during the tka procedure (inside the patient); therefore, the surgeon ¿used a saw blade instead¿ and was able to complete procedure ¿without delay or harm¿.Reportedly, the angel wing ¿had been warped¿.Based on the information provided, further patient impact would not be anticipated, as the procedure was reportedly completed with an alternate device, without patient injury/harm or surgical delay.No further medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY RESECTION CHECK
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11094903
Report Number1020279-2020-08010
Device Sequence Number3698392
Product Code HRY
UDI-Device Identifier03596010556738
UDI-Public03596010556738
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74012431
Device Catalogue Number74012431
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Initial Date Received by Manufacturer 12/07/2020
Supplement Date Received by Manufacturer07/12/2021
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date07/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-