Visual inspection: the tulip was returned with the fractured shank bulb still attached inside of it.The rest of the screw shank was not returned.The shank was fractured at the bulb.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.From the ifu: delayed union or nonunion: internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs.In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture.The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant.If a nonunion develops or if the implants loosen, bend or break, the device(s) should be revised or removed immediately before serious injury occurs.The screw was implanted for over a year and the patient had a non union.Additionally, the bone quality was poor due to a patient infection that could cause instability in the construct.The likely cause is multifactorial with these factors likely contributing to the event.
|