| Model Number |
DBEC-125 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).System instructions for use manual states that slow flow and no reflow is a possible adverse event which can occur with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Three treatment passes on low speed with a diamondback coronary orbital atherectomy device (oad) from the proximal to mid left anterior descending artery (lad).The treatment times were long.Another treatment was performed at high speed in the proximal lad, and the oad was removed.Imaging was performed, and slow flow was identified in the distal lad.The patient complained of chest pain.Nitroglycerin and adenosine were administered; normal flow was restored, and the patient's chest pain was resolved thereafter.The procedure was completed with angioplasty and stent placement.The patient was stable and was scheduled for discharge as planned.
|
| |
|
Search Alerts/Recalls
|