It was reported that during procedure, tan/fan perc drill guide was bent and not able to connect properly.No harm or injury reported on patient.Procedure was finished with the same device.No delay reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device was manufactured in 2013 and shows signs of extensive use.A functional evaluation cannot confirm the stated failure mode, due to the paired mating part mechanism being stuck together.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|