SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71362293
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reference: (b)(4).
 
Event or Problem Description
It was reported that, the doctor was using the straight hex driver on an acetabular screw & the hex spun out of the screw.After inspection it was noticed that the tip of the straight hex driver was worn down.The procedure was completed with a minimal delay using a smith-nephew back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SCREWDRIVER
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11095896
Report Number1020279-2020-08042
Device Sequence Number11687147
Product Code HRY
UDI-Device Identifier03596010454188
UDI-Public03596010454188
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362293
Device Catalogue Number71362293
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/08/2020
Supplement Date Received by Manufacturer04/20/2021
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date04/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
-
-