DEPUY SPINE INC VIPER 2 SYSTEM SCREW EXTENSION, OPEN; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Model Number |
286725300 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, the cage broke during impaction into hard bone where the space was extremely collapsed.Screws were later inserted to back up the interbody fusion and the screwdrivers stripped during final tightening attempting to back out the screws and adjust the height.No fragments were generated.The cage will stay implanted.The surgeon packed the rest of the space with bone.Procedure was successfully completed with no surgical delay.There was no patient consequences.Concomitant device reported: unknown impaction instruments (part# unknown, lot# unknown, quantity unknown); unknown insertion instruments (part# unknown, lot# unknown, quantity unknown); cougar ls 18x14x50 15-deg lord (part # 189150614, lot #unknown, quantity 1).This report is for one (1) viper 2 system screw extension, open.This is report 7 of 8 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D11: concomitant product added to complaint.Investigation summary: background: 12/07/2020: updated ed: it was reported that on an unknown date, the cage broke during impaction into hard bone where the space was extremely collapsed.Screws were later inserted to back up the interbody fusion and the screwdrivers stripped while final tightening and also while attempting to back out the screws to adjust the height.No fragments were generated.The kh (cage) which fractured was not removed from the patient.It was kept in place as it was determined to be still structurally sound despite having fractured.Procedure was successfully completed with no surgical delay.There was no patient consequences.Concomitant device reported: unknown impaction instruments (part# unknown, lot# unknown, quantity unknown ).Unknown insertion instruments (part# unknown, lot# unknown, quantity unknown ).This complaint involves five (5) devices.Investigation flow: damage.Visual inspection: the viper2 final tightener driver (part #: 286725300, lot #: 0510mi ) was received at us cq.Visual inspection of the compliant device showed that the distal tip was worn/damaged.The worn condition of the distal tip is consistent as an end of life indicator for the device.No other issues were identified during inspection.The received condition is consistent with the complaint condition thus the complaint is confirmed.Conclusion: after a visual inspection per guidance provided in windchill document # 0000277191, it is determined that the device is worn from repeated use and servicing therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities device history lot: a review of the receiving inspection (ri) for viper2 screw extension, open was conducted identifying that lot number 0510mi was released in a single batch.Batch1: lot qty of (b)(4) units were released on 07 jul 2010 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device history batch: null.Device history review: the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- the viper2 final tightener driver (part #: 286725300, lot #: 0510mi ) was received at us cq.Visual inspection of the compliant device showed that the distal tip was worn/damaged.The worn condition of the distal tip is consistent as an end of life indicator for the device.No other issues were identified during inspection.The received condition is consistent with the complaint condition thus the complaint is confirmed.Conclusion: after a visual inspection per guidance provide, it is determined that the device is worn from repeated use and servicing therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities device history lot : a review of the receiving inspection (ri) for viper2 screw extension, open was conducted identifying that lot number 0510mi was released in a single batch.Batch1: lot qty of units were released on (b)(6) 2010 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device history review : the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|