• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53 - CARDIOSAVE HYBRID DOMESTIC
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).During a pm service by a getinge field service engineer (fse), the pim failed during test.The fse replaced the pneumatic module assembly, then completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) had pim failure.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h6(evaluation method codes), h10, h11.Corrected fields: b5, h6(device codes).
 
Event Description
It was reported that during a preventive maintenance (pm) performed by a getinge field service engineer (fse), the pim of the cardiosave intra-aortic balloon pump (iabp) failed the pim testing.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID DOMESTIC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11096217
MDR Text Key224598813
Report Number2249723-2020-02246
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53 - CARDIOSAVE HYBRID DOMESTIC
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/07/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received01/15/2021
Supplement Dates FDA Received02/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-