DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 - CARDIOSAVE HYBRID DOMESTIC |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).During a pm service by a getinge field service engineer (fse), the pim failed during test.The fse replaced the pneumatic module assembly, then completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) had pim failure.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h6(evaluation method codes), h10, h11.Corrected fields: b5, h6(device codes).
|
|
Event Description
|
It was reported that during a preventive maintenance (pm) performed by a getinge field service engineer (fse), the pim of the cardiosave intra-aortic balloon pump (iabp) failed the pim testing.There was no patient involvement, and no adverse event reported.
|
|