• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIGRID BULK 6X7CM CASE 2100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIGRID BULK 6X7CM CASE 2100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000376
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during treatment with opsite film, the dressing fell off.It is unknown how the procedure was finished and if there was a delay.No other complications were reported.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the batch manufacturing records could not be performed as no lot number was provided, however, there are no indications to suggest that the device did not meet specifications upon release into distribution.The complaint history file contains no further instances/ related events of the reported event.The device was used for treatment.No samples were returned for analysis.Low adhesion may be caused by adhesion levels of the raw material used to make the film of this dressing.There are various controls in place during the manufacturing process of both the raw material and the dressing to identify when the adhesion is not in acceptable range, as per the product specification.There are also regular tests carried out to monitor the finished product.As with all adhesive products, the skin site must be thoroughly cleaned and dried prior to application.If there is any moisture on the skin surrounding the catheter, the adhesive performance of this dressing may be compromised.If the dressing becomes wet whilst applied, it may have to be removed and a new dressing applied.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIGRID BULK 6X7CM CASE 2100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11096320
MDR Text Key224415384
Report Number8043484-2020-04408
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223415402
UDI-Public05000223415402
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2021
Device Catalogue Number66000376
Device Lot Number1357461928
Date Manufacturer Received02/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-