DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The getinge field service engineer (fse) that encountered the issue replaced the pneumatic module assembly and performed 30 psi calibration procedure.The fse found the fiber optic connector was damaged and replaced the front end assembly.Pm was completed during the same service visit.Full calibration, functional, and safety checks were performed which passed to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the x1 transducer of the cardiosave intra-aortic balloon pump (iabp) would not calibrate.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the 30 psi calibration could not be completed and the x1 transducer of the cardiosave intra-aortic balloon pump (iabp) would not calibrate.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Corrected fields: b5, d1, e1 (event site name, event site address, event site postal code, event site telephone).
|
|