• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The getinge field service engineer (fse) that encountered the issue replaced the pneumatic module assembly and performed 30 psi calibration procedure.The fse found the fiber optic connector was damaged and replaced the front end assembly.Pm was completed during the same service visit.Full calibration, functional, and safety checks were performed which passed to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the x1 transducer of the cardiosave intra-aortic balloon pump (iabp) would not calibrate.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the 30 psi calibration could not be completed and the x1 transducer of the cardiosave intra-aortic balloon pump (iabp) would not calibrate.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Corrected fields: b5, d1, e1 (event site name, event site address, event site postal code, event site telephone).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11096412
MDR Text Key224598683
Report Number2249723-2020-02247
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-