Model Number 3577
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
 
Event or Problem Description
The customer reported the device is providing low readings and the plastic guard has broken off.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned device was evaluated.During visual inspection, the bend relief sleeve was found to be damaged and the latch is missing.The device passed functional testing and the module maintains communication and measurement with cable bending.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit had no product performance issues related to spo2 and pulse rate.The device functions as designed regarding accuracy and measurements.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
 
Event or Problem Description
The customer reported the device is providing low readings and the plastic guard has broken off.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
USPO2 PULSE OXIMETRY CABLE KIT
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key11096515
Report Number3011353843-2020-00214
Device Sequence Number700629
Product Code DQA
UDI-Device Identifier00843997007481
UDI-Public00843997007481
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K121914
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3577
Device Catalogue Number3577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/08/2020
Supplement Date Received by Manufacturer12/31/2020
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date01/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-