| Model Number |
3577 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the device is providing low readings and the plastic guard has broken off.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: other, other text: the returned device was evaluated.During visual inspection, the bend relief sleeve was found to be damaged and the latch is missing.The device passed functional testing and the module maintains communication and measurement with cable bending.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit had no product performance issues related to spo2 and pulse rate.The device functions as designed regarding accuracy and measurements.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
|
| |
|
Event or Problem Description
|
|
The customer reported the device is providing low readings and the plastic guard has broken off.No patient impact or consequences were reported.
|
| |
|