| Model Number |
168-510 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: patient's concern with the product, deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "left side deflation and exchange from textured to smooth breast implants due to the patient¿s concern with the product".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: crease fold, wear abrasion and opening on anterior.Leak test and microscopic analysis was performed which identified: crease smooth opening.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: an opening crease smooth on anterior assessed as fold flaw opening.
|
| |
|