MAUDE Adverse Event Report: ARTHROCARE CORP. MICROBLATOR 30 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. MICROBLATOR 30 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number AC4050-01
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that during setup when the wand was connected it did not perform any function.The procedure was completed with a competitor device and no delay or further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use (ifu) was reviewed and contains warnings and precautionary measures related to proper use of the device.A review of risk management files found that the reported failure was documented appropriately.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: (1) misalignment or connecting the wand to the unintended port which could have caused damage to the pin, thus not performing any functions.(2) there may have been an issue with one of the concomitant devices in use at the time of this complaint event.No containment or corrective actions are recommended at this time.There are no indications to suggest the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows the block connector has bent pins.A functional evaluation could not be conducted due to the connector having bent pins.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The complaint history review found further instances of the reported event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows the block connector has bent pins.A functional evaluation could not be conducted due to the connector having bent pins.The bent pins prevent the wand from connecting to the controller, therefore no testing can be completed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution the instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was not confirmed and the root cause could not be determined as the condition in which the device was received did not allow for evaluation of the reported complaint.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROBLATOR 30 ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11097395
Report Number3006524618-2020-01169
Device Sequence Number4422394
Product Code GEI
UDI-Device Identifier00817470003123
UDI-Public00817470003123
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/10/2022
Device Model NumberAC4050-01
Device Catalogue NumberAC4050-01
Device Lot Number2037547
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2021
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer01/28/2021
06/18/2021
10/08/2021
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date02/08/2021
07/07/2021
10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-