A facility reported that the certas inlin valve was implanted on (b)(6) 2020 to a patient via v-p shunt on unknown date with setting 5.However, on (b)(6) 2020, ventricular enlargement was observed under mri without any clinical symptoms.The physician suspected valve obstruction.Therefore, the valve will be replaced to a new one (828800) on (b)(6) 2020.
|
Unique device identifier (udi) : (b)(4).The certas valve was returned for evaluation: failure analysis - the position of the cam when valve was received was at setting 1.The valve was visually inspected; the needle guard was raised and needle holes in the needle chamber were noted.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.The needle chamber was dismantled; the needle guard was bent, and glue traces were noted on the needle guard and the silicone housing base.The root cause for the problem reported by the customer is due to the raised needle guard this is probably due to wrong handling as noted in the "ifu" : do not fold or bend the valve, folding or bending might cause rupture of the silicone housing, needle guard disc dislodgement or occlusion of the fluid pathway.
|