| Model Number |
2800-55 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An evaluation of the acl top 750 cts by the software development team determined that the barcode was not misread due to a software defect or instrument malfunction.The software development team indicated that a barcode may be misread due to human factors such as when a label is twisted/tilted hiding part of the barcode or if the printed labels are of poor quality.It is likely that the barcode was not placed properly and therefore was misread when placed on the sample rack.Acl top family 50 series operator's manual, in chapter 4 indicates the proper placement in "bar code label placement".No remedial action required.
|
| |
|
Event or Problem Description
|
|
The complaint stated that there was a malfunction of the barcode reader on the acl top 750 cts instrument.The instrument barcode reader misread a patient sample identification number and the sample results were assigned to another patient.The results were not transmitted to the laboratory information system (lis) and there was no patient impact as a result of this event.
|
| |
|