BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.25/13; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401735
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a very calcified lesion in the rca.After pre-dilatation the target lesion could not be crossed and the orsiro came off in the ostium of the rca.Eventually the stent could be retrieved by using a balloon catheter.An athrectomy was then performed.
 
Additional Manufacturer Narrative
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the balloon has clearly been inflated and contrast medium was found inside of the balloon.Stent imprints are visible on the balloon surface, indicating that the stent was initially crimped centered on the balloon.The stent was returned separately, unprotected and is severely deformed over its entire length.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.Considering the physicians description, the root cause for the complaint event is most likely related to the patients anatomy (i.E.Severely calcified lesion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 2.25/13
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11098013
Report Number1028232-2020-05715
Device Sequence Number4314552
Product Code NIQ
UDI-Device Identifier07640130438903
UDI-Public07640130438903
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/05/2021
Device Model Number401735
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10194428
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Received by Manufacturer 12/18/2020
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-