| Model Number |
BXA093901A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, the patient was being implanted with a gore® viabahn® vbx balloon expandable endoprosthesis (vbx) to treat stenosis of the brachiocephalic artery.An 8 fr silk road sheath was used during the procedure.It was reported the vbx was advanced and implanted with no issue.However, during withdrawal of the catheter resistance was met.The catheter was able to be withdrawn through the sheath.It was noted the balloon had become separated from the catheter and was located within the portion of the sheath in the patient.The sheath and balloon were successfully withdrawn from the patient together.The patient tolerated the procured.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated h6 investigation conclusion - 4315 cause not established.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.
|
| |
|