Model Number GF-210RA
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported that their multigas unit is displaying a "gas device failure" error when it's in use.The customer also reported that the device pump is not turning on.No harm or injury was reported.They will be sending this unit in for an exchange.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that their multigas unit is displaying a "gas device failure" error when it's in use.
 
Event Description
The customer reported that the multigas unit was displaying a "gas device failure" error message while in patient use.The customer also reported that the device pump was not turning on.
 
Manufacturer Narrative
Details of complaint: the customer reported that the multigas unit was displaying a "gas device failure" error message while in patient use.The customer also reported that the device pump was not turning on.They sent this unit in for an exchange.No patient harm was reported.Service requested / performed: exchange/evaluation.Investigation summary: the root cause is determined to be component failure.The sensor needed replacing.The sensor is a replaceable component, where the longevity is dependent upon the following factors: instrument and parts must undergo regular maintenance inspection at least every 6 months.If stored for extended periods without being used, make sure prior to operation that the instrument is in perfect operating condition.Water trap should be replaced every 4 weeks or as indicated on the device.Ensuring the environment is optimal as described in the operator's manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GF-210RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key11098835
MDR Text Key226271368
Report Number8030229-2020-00803
Device Sequence Number1
Product Code CCK
UDI-Device Identifier04931921106891
UDI-Public04931921106891
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110594
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-210RA
Device Catalogue NumberGF-210RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-