Model Number Z9002
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Vitrectomy (2643); No Code Available (3191)
Date of Event 06/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided.Best estimate of date of event is around (b)(6) 2020 as the timeframe provided for symptom onset is six months ago.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient¿s left eye due to lens dislocation.It was replaced with a non-johnson and johnson iol but same diopter power.A vitrectomy was performed.No other information was provided.
 
Additional Manufacturer Narrative
Section d10.Device available for evaluation? yes.Returned to manufacturer on: 1/6/2021.Section h3.Device returned to manufacturer? yes.Device evaluation: returned lens was received, submerged in an unknown solution inside a specimen cup.Additional documentation was received as well.The lens was cleaned and visual inspection under magnification revealed, that the lens was returned cut in half with a bent haptic.Which may have occurred, when the lens was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed.And no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed, that there was no discrepancy found during the review.The product was manufactured and released according to specifications.A search revealed, that two additional complaints were received from this production order, but no product malfunction or discrepancy could be identified.And no escalation is required.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key11099305
Report Number2648035-2020-00978
Device Sequence Number3630662
Product Code HQL
UDI-Device Identifier05050474530676
UDI-Public(01)05050474530676(17)181227
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/27/2013
Device Model NumberZ9002
Device Catalogue NumberZ900200190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2021
Initial Date Received by Manufacturer 12/03/2020
Supplement Date Received by Manufacturer01/31/2021
Initial Report FDA Received Date12/30/2020
Supplement Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-