SMITH & NEPHEW, INC. LAG SCREW DRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674067
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/09/2020
Event Type  Injury  
Manufacturer Narrative
Complaint reference: (b)(4).
 
Event Description
It was reported that during surgery inside the patient the lag screw driver tip broke off.No delay.The procedure was finished with the same device.
 
Manufacturer Narrative
The device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken and has signs of wear and tear from use.The device was manufactured in 2019.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints at this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11099646
MDR Text Key224520280
Report Number1020279-2020-08089
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560605
UDI-Public03596010560605
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674067
Device Catalogue Number71674067
Device Lot Number19DCW3016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Date Manufacturer Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-