WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 70468580
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
(b)(6) 2020, mpxr report (b)(4), (rep, hcp,for): information was received from a health care provider via a manufacturer representative regarding a patient with tumor on right side in s1 for open procedure at t11 - ilium and s1 / l5 spinal therapy.It was reported that the rod and screws was broken on left side near and in s1.There was pain symptoms reported for patient involved in this event.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3:part # 70468580; lot # 0269719w visual examination revealed the entire bone screw has broken at the head.Optical examination of the area of fracture initiation did not identify a pre-existing surface defect that could contribute to crack propagation.Some fracture surface smearing is noted.Microscopic examination of the fracture surface identified a fairly flat fracture surface with progressive striations consistent with cyclic fatigue.The threads of the broken lower portion of the screw have been damaged, possibly from the removal process.This type of damage is consistent with cyclic fatigue followed by overload once the fatigue had weakened the material.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key11099706
Report Number1030489-2020-01879
Device Sequence Number19842862
Product Code KWP
UDI-Device Identifier00613994158222
UDI-Public00613994158222
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K050439
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number70468580
Device Catalogue Number70468580
Device Lot Number0269719W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2020
Supplement Date Received by Manufacturer03/18/2021
Initial Report FDA Received Date12/31/2020
Supplement Report FDA Received Date03/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-