MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1714K 630G BLACK MMOL CANADA; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1714K
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that the insulin pump had critical pump error alarm.Customer reported that the insulin pump was started alarming and display was flashing.Customer stated that the insulin pump had open book image.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with critical pump error (open book) due to moisture damage electronics.Unable to performed download due to moisture damage.Unable to performed displacement test, rewind test, prime or seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self test due to critical pump error.Device received with moisture damage noted on motor, vibrator motor or battery tube assembly.Device received with cracked battery tube threads and cracked case corner of belt clip rails.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1714K 630G BLACK MMOL CANADA
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key11100009
MDR Text Key225731138
Report Number2032227-2020-222758
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000315702
UDI-Public(01)000000763000315702
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1714K
Device Catalogue NumberMMT-1714K
Device Lot NumberHG4L3WV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2021
Date Manufacturer Received02/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-