Model Number 799-EGB2-01-01
Medical Device Problem Codes Calibration Problem (2890); Pressure Problem (3012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to treat a (b)(6)-year-old male patient, the device displayed "runtime calibration failure - 1051" and "airway pressure sensing failure 1052" messages.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
The customer's report was observed during review of the device activity logs.However, the device was put through extensive testing including full functional testing and power cycling without duplicating a malfunction to the device.Internal and visual inspections found no discrepancies.The smart pneumatic module board was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Common Device Name
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key11100196
Report Number1220908-2020-04283
Device Sequence Number14388503
Product Code CBK
UDI-Device Identifier00847946026592
UDI-Public00847946026592
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K111473
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number799-EGB2-01-01
Device Catalogue Number799-EGB2-01-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/12/2020
Supplement Date Received by Manufacturer12/12/2020
Initial Report FDA Received Date12/31/2020
Supplement Report FDA Received Date01/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age54 YR
-
-