Model Number 21-7302-24
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2020
Event Type  malfunction  
Event Description
Information was received indicating that when filling the smiths medical cadd cassette reservoir, the bag has been breaking in the cassette, leaking the medication.It was reported that a syringe attached at the end was used to add drug and it started to leak out of the cassette.No patient consequences were reported.No adverse effects were reported.
 
Manufacturer Narrative
Other, other text: once cassette was returned.Leak testing was performed and the unit was found to be leaking on the joint of the tube.The bag sealer machine id# bfm-2-1-001 was causing the weak seal condition on the joint of the tube with the bag, the generator of the equipment was identified.Procedure not followed: the two forms related to the procedures pm-1011 and pm-1026, for to register units with leak were not filling up, even that was identified the equipment malfunction, as well the method to test the units in both process was not followed as is required, forcing the equipment to release parts with potential leak condition.No corrective actions have been identified at this time; our investigation remains in progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
MDR Report Key11100317
MDR Text Key224528146
Report Number3012307300-2020-12915
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number4046848
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-