Model Number 8300
Device Problems Break (1069); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The customer reported problem was confirmed.A review of the device history record in sap for sn (b)(4) was performed from the date of the manufacture to date of the release of product, which confirmed that this device was not involved in a production failure, and product was returned for servicing which correlates to the customer reported issue.A review of the complaint history record in the trackwise was performed for the sn (b)(4) which confirmed no similar complaints with the same or related failure mode.The customer stated that there was no patient involvement.
 
Event Description
Doesnt recognize disposables.Oridion board- comm failure.Iui- isolation rib damage -(b)(4).There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8300 ALARIS ETCO2 MODULE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
madhavan ezhumalai
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key11101249
MDR Text Key228736356
Report Number2016493-2020-86083
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10885403830013
UDI-Public10885403830013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8300
Device Catalogue Number8300
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-