ZIMMER BIOMET, INC. 5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD; PROSTHESIS, TRAUMA Back to Search Results
Catalog Number 47248405050
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: foreign - event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported approximately three (3) weeks ago, as the cardboard outer packaging was being removed from the znn screw, the inner plastic barrier was compromised.The box was opened at the exact place it is perforated, but the inner plastic was broken at the same time.No surgery occurred and no patient involvement.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the provided picture identified that the extremity of the sterile cavity is bent along the folding line of the cardboard box and resting within the folded carton box portion.Medical records were not provided.Dhr was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event it likely to be damage during transit.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned product found the outer carton and the outer sterile blister are damaged.Due to the damage to the sterile blister, the sterility of the product has been breached.The root cause of the event remains the same as previously reported.The root cause of the reported event is likely to be damage during transit.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 MM DIAMETER CORTICAL SCREW - RED FIXED ANGLE 3.5 MM HEX HEAD
Common Device Name
PROSTHESIS, TRAUMA
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11101468
Report Number0001822565-2020-04264
Device Sequence Number16964740
Product Code HSB
UDI-Device Identifier00889024094161
UDI-Public(01)00889024094161
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K082770
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number47248405050
Device Lot Number64738657
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/10/2020
Supplement Date Received by Manufacturer01/15/2021
06/04/2021
Initial Report FDA Received Date12/31/2020
Supplement Report FDA Received Date01/15/2021
06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
-
-