• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL 15 DEG TAPER 3.5P 3MM-4MM; DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL 15 DEG TAPER 3.5P 3MM-4MM; DENTAL ABUTMENT Back to Search Results
Model Number 15AT334
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Weight unknown / not provided.Fax number unknown / not provided.
 
Event or Problem Description
It was reported that the abutment did not seat to the implant.Doctor placed another abutment to the implant and finished the procedure.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).One 15 deg taper 3.5p 3mm-4mm (15at334) was returned for investigation.Visual inspection of the as returned product identified signs of wear from use about the abutment body, and some debris on the threads.Functional testing of the returned device notes that the abutment assembled and fully seated with a mating implant.No pre-existing conditions were noted on the per.The reported product was located on an unknown tooth site and was removed immediately.Review of appropriate documentation: documents reviewed: 4894 rev 6 - 08/19; adverse effects; page 4 dhr review was completed for the subject lot number 2020060512.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2020060512) for similar event and no other complaint was identified.January post market trending was reviewed and there were no actionable events or corrective actions for the reported event (component does not seat) or product (15at334).Therefore, based on the available information, device malfunction has not occurred and the reported event was unconfirmed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4) the following sections have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h10: additional narrative this event has been reassessed as not reportable, this malfunction is not likely to cause or contribute to a serious injury or death if it were to recur.The most likely outcome would be customer / patient inconvenience resulting in an additional dental appointment to refit the abutment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
15 DEG TAPER 3.5P 3MM-4MM
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key11102258
Report Number0002023141-2020-02443
Device Sequence Number3781744
Product Code NHA
UDI-Device Identifier00889024012059
UDI-Public(01)00889024012059(17)250629(10)2020060512(241)15AT334
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K111853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/29/2025
Device Model Number15AT334
Device Catalogue Number15AT334
Device Lot Number2020060512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/09/2020
Supplement Date Received by Manufacturer03/18/2021
05/04/2021
Initial Report FDA Received Date12/31/2020
Supplement Report FDA Received Date04/06/2021
05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-