It was reported that, during a cori tka live broadcast procedure, when the doctor lifted his foot off the black pedal when he was done painting/surface mapping the femur and tibia, the system kept collecting points as he lifted the probe off the bone.They had to hit the "stop collection" button on the 24 inch touchscreen both times to stop the data collection to proceed.No other complications were reported.
|
H3, h6: the real intelligence cori, pn: rob10024, sn: (b)(6) used for treatment was not returned to the designated complaint unit for evaluation, and the cori log files and/or case files were not provided.Therefore visual and functional inspections and log/case file assessments could not be performed, and the reported problem could not be confirmed.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports, this issue will continue to be monitored.Although the reported event could not be confirmed, the most likely cause of the point collection occurring after the foot pedal is released is a known software bug that is under investigation by the software engineering team.No containment or correction is required at this time.
|