• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2020
Event Type  malfunction  
Event Description
It was reported that, during a cori tka live broadcast procedure, when the doctor lifted his foot off the black pedal when he was done painting/surface mapping the femur and tibia, the system kept collecting points as he lifted the probe off the bone.They had to hit the "stop collection" button on the 24 inch touchscreen both times to stop the data collection to proceed.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the real intelligence cori, pn: rob10024, sn: (b)(6) used for treatment was not returned to the designated complaint unit for evaluation, and the cori log files and/or case files were not provided.Therefore visual and functional inspections and log/case file assessments could not be performed, and the reported problem could not be confirmed.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports, this issue will continue to be monitored.Although the reported event could not be confirmed, the most likely cause of the point collection occurring after the foot pedal is released is a known software bug that is under investigation by the software engineering team.No containment or correction is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11102481
MDR Text Key224686858
Report Number3010266064-2020-02157
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Date Manufacturer Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ROB20000, SERIAL: (B)(6).
-
-