| Model Number |
ZCU150 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.If implanted, give date: not applicable, the lens was removed/replaced within the initial procedure.If explanted, give date: not applicable, the lens was removed/replaced within the initial procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the toric intraocular lens (iol) back haptic did not unfold when inserting into the patient's eye.The procedure was completed with a back up lens.There was no additional surgical intervention required and patient outcome was not provided.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product evaluation was not performed because the product has not been received.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|