BELMONT MEDICAL TECHNOLOGIES THE BELMONT RAPID INFUSER; THERMAL INFUSION FLUID WARMER Back to Search Results
Model Number FMS 2000
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The rapid infuser, fms 2000 and associated disposable set have been returned to belmont for investigation; evaluation of the unit is in process.When the rapid infuser detects a situation that is compromising effective infusing, the system stops pumping and heating, closes off the line to the patient, sounds an audible alarm, and displays an alarm message with instructions for corrective measure.In the event of an "over temperature" alarm, the rapid infuser exhibits the following alarm message: "infusate over temperature.Discard disposable and blood.Restart system with a new disposable.Service machine if error persists." the operator's manual also provides possible conditions and additional recommended operator actions.There is not enough information available at this time to determine the root cause of the reported over temperature alarm.It was reported that there was no injury to the patient.The manufacturing records for this serial number were reviewed and no anomalies were identified.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Belmont's distributor received a report from the user facility that a burning odor was noted after infusing 200ml of prbcs and warmed plasma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE BELMONT RAPID INFUSER
Common Device Name
THERMAL INFUSION FLUID WARMER
Manufacturer (Section D)
BELMONT MEDICAL TECHNOLOGIES
780 boston road
billerica MA 01821
MDR Report Key11103306
Report Number1219702-2020-00126
Device Sequence Number14313766
Product Code LGZ
UDI-Device Identifier10896128002012
UDI-Public(01)10896128002012
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K141654
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFMS 2000
Device Catalogue Number903-00001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2020
Initial Report FDA Received Date12/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-