The customer did not return the suspected contour next test strips for evaluation.Therefore, the device history record was reviewed for the suspected lot of test strips and no manufacturing anomalies were found.Additionally, the labeling scans were reviewed for the following lot #s 0cpeg12a, 0cpeg12b, 0cpeg12c, and 0cpeg12d based on the raw lot # 0cpeg12, and no labeling issues were identified.During the packaging, a vision inspection test is performed which would remove any vials if there was problem identified with the label.
|