| Model Number |
E053 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Codes |
No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been received for analysis.This report would be updated should further information be provided from the analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited premature battery depletion (pbd) behavior.The battery appeared to be depleting more quickly than expected.Device replacement was recommended.Eventually, this device was surgically explanted, replaced and returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient code 3191 captures the reportable event of surgery.The product has been received for analysis.This report would be updated should further information be provided from the analysis.The returned implantable cardioverter defibrillator was analyzed.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device battery.Low voltage capacitors are used in the high voltage charging operation in order to facilitate fast charge times.The behavior of these capacitors resulted in a high current drain, which was depleting the device battery faster than normal.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited premature battery depletion (pbd) behavior.The battery appeared to be depleting more quickly than expected.Device replacement was recommended.Eventually, this device was surgically explanted, replaced and returned for analysis.No additional adverse patient effects were reported.
|
| |
|