Model Number E053
Medical Device Problem Codes Premature Discharge of Battery (1057); Noise, Audible (3273)
Health Effect - Clinical Codes No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis, should pertinent information be provided.(b)(4).
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) emitted beeping tones and recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.The device was explanted and returned for analysis.No additional adverse patient effects.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) emitted beeping tones and recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.The device was explanted and returned for analysis.No additional adverse patient effects.
 
Additional Manufacturer Narrative
The returned implantable cardioverter defibrillator was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device battery.Low voltage capacitors are used in the high voltage charging operation in order to facilitate fast charge times.The behavior of these capacitors resulted in a high current drain, which was depleting the device battery faster than normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key11104746
Report Number2124215-2020-26187
Device Sequence Number14433861
Product Code LWP
UDI-Device Identifier00802526480706
UDI-Public00802526480706
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S235
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/09/2014
Device Model NumberE053
Device Catalogue NumberE053
Device Lot Number100845
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Initial Date Received by Manufacturer 11/12/2020
Supplement Date Received by Manufacturer02/02/2021
Initial Report FDA Received Date01/02/2021
Supplement Report FDA Received Date02/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age44 YR
-
-